Controller Page — User Guide

Open-LIFU Test App

Docs Navigation: README | Launch Options | Transmitter | Console | Verification | Settings | Support

In This Page


Overview

Back to top

The Controller page is the main engineering surface for configuring and running an arbitrary sonication. It exposes every parameter the firmware accepts, lets you load and save full solutions to JSON, and drives the device through its READY → RUNNING → READY cycle.

The Controller has limited solution generation capabilities via it's input fields, but does not have the full flexibility of the OpenLIFU application. For more complex sonication patterns, users should generate a JSON solution using the SDK and load it into the Controller. Solutions loaded from JSON disable the input fields, and respect the parameters as authoritative — the Controller becomes a live dashboard for monitoring and running the solution, but not editing it.

Navigate to the Controller page by clicking the controller icon in the left sidebar.

Controller tab — full layout


Inputs Panel

Back to top

Voltage

Field Units Notes
Voltage per Rail (+/-) V Peak-to-peak is double this value

Sequence Settings

Field Units Notes
Trigger Mode Single (one train), Continuous (repeated indefinitely), Sequence (fixed count)
Pulse Interval ms Time between pulse starts
Pulse Count Pulses per train (disabled in Single mode)
Pulse Train Interval s Time between train starts (disabled in Single mode)
Pulse Train Count Number of trains (Sequence mode only)

When using Trigger Mode "Single", the Pulse Train Interval and Pulse Train Count fields are disabled, and the device will emit a single train of pulses when Start is pressed. In "Continuous" mode, the device will emit trains of pulses indefinitely until Stop is pressed; the Pulse Train Interval field is enabled but ignored, and the Pulse Train Count field is disabled. In "Sequence" mode, the device will emit a fixed number of pulse trains as specified by the Pulse Train Count field; both the Pulse Train Interval and Pulse Train Count fields are enabled.

Pulse Settings

Field Units Notes
Frequency kHz Stored/sent as Hz internally
Duration µs Pulse-on time within each pulse

Focus Point

Field Units Notes
X / Y mm Lateral offset, typically 0 / 0
Z mm Steering depth from the transducer face

Field colour conventions

  • White — pristine value, never sent
  • Green — value confirmed by the device after a successful Configure
  • Grey — section is read-only (a solution was loaded from a JSON file, or the device is currently running)
  • Orange — value is being directly edited but not yet committed to the device

Once the device has been configured at least once on a manually-entered solution, edits to individual fields are committed straight to the device when you press Enter or tab away — no need to re-Configure for each change.

Solution Load/Save controls

At the bottom of the left column are controls for loading and saving sonication solutions as JSON files. This is the only way to interact with the more complex parameters like arbitrary delays and element apodizations.

Selecting Load will pull up the Load Solution Dialog: Load Solution Popup

Control Action
Preset dropdown Pick a JSON solution from the presets solution folder
Load from File Open a file picker for an arbitrary .json solution
Cancel Close the dialog without loading a solution
Load Load the selected solution, close the dialog

Loading a solution from disk locks the input fields (read-only grey) because the JSON is treated as authoritative.

Loaded Solution

Selecting Save will open the Save Solution Dialog: Save Solution Popup

Control Action
Solution_id Identifer for the solution, used as the filename when saving
Solution Name Human-friendly name for the solution, stored in the JSON metadata
File Location Directory to save the JSON solution to
Browse Open a directory picker to select the File Location
Save as Default Save the solution as the default parameters to use when the app is opened
Cancel Close the dialog without saving
Save Save the solution to file and close the dialog

Clicking Edit Solution after loading a solution from JSON will drop the loaded solution, unlock the fields, and allow you to edit parameters manually. This is a one-way operation — once you click Edit Solution, the Controller will no longer reference the loaded JSON for any parameters, and any changes you make in the UI will not be reflected in the JSON. If you want to go back to the loaded JSON parameters, you can either re-load the solution from file or click Reset to drop all parameters and return to the default state.


Plot Panel

Back to top

The plot shows the pulse to be generated, the envelope of the pulse train, and a representation of the time delays applied to each element. The depiction of the time delays uses a fixed approximation of the element position based on whether or not there are 1 or 2 modules connected.


Status and Actions Panel

Back to top

Connection / status header

Indicator Meaning
System State Mirrors LIFUConnector.state: Disconnected / Connected / Ready / Running
Modules Number of connected TX modules
HV Mode AUTO / ON / OFF / WHILE_RUNNING
TX LED Red = disconnected · Dark green = connected · Green = ready · Blue = running
Temp [...] Per-module TX driver temperature in °C
HV LED Red = disconnected · Dark green = connected · Green = ready · Blue = HV rail energised
Rails +x.xx / -x.xx V Most recent monitored HV rail readings
Progress Bar See below

Action buttons

Button Behaviour
Configure Validate the current parameters and push them to the firmware. Required at least once before Start can be enabled.
Reset Drop the loaded solution, unlock fields, drop HV (if AUTO), and return to CONNECTED
Start Begin sonication. Disabled until the device is in READY
Stop Abort the current run

HV enable mode

Mode Behaviour
AUTO (default) HV held on whenever the TX has a valid solution loaded
ON HV held on continuously
OFF HV held off; Start is blocked
WHILE_RUNNING HV energised only while sonication is actually running

Sonication progress

A progress bar at the bottom of the right column tracks the current run.

State Colour Text
Idle Grey (empty)
Running Blue RUNNING i/N (Sequence) or RUNNING i (Continuous)
Finished Green FINISHED N/N
Stopped Orange STOPPED

Progress updates are driven by unsolicited STATUS frames from the firmware, so the bar advances even when the app is not actively polling.


Typical workflow

Back to top

  1. Connect the Console and Transmitter — the TX and HV LEDs go green.
  2. Either pick a preset, load a solution from disk, or fill in the parameter fields manually.
  3. Click Configure. Fields turn green; the plot refreshes; the System State moves to Ready.
  4. Click Start. The progress bar begins to fill and the system state flips to Running.
  5. Wait for FINISHED (if using Single or Sequence mode), or click Stop to abort.
  6. Edit any field and press Enter — the new value is pushed live without a full re-Configure (assuming you have not loaded a solution from file).